Customization: | Available |
---|---|
After-sales Service: | Yes |
Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Equipment use and Introduction
For the determination of medical protective clothing products in the dry state of the number of flocculation and cleanliness - particulate matter index .
Protective clothing dry state flocculation tester is used to determine the number of flocculation and cleanliness of medical protective clothing products in the dry state - particulate matter index.
Technical parameters
1,Disk diameter: Φ82.8mm
2,Reciprocating movement frequency: 60 times/min
3,Rotation angle: 180 degrees/time (clockwise and counterclockwise alternately)
4,Maximum starting distance between two discs: (188±2)mm
5,Straight dynamic range of dynamic disc: (120±2)mm
6,Size of antistatic test chamber: 300mm*300mm*300mm
7,The diameter of air collector sampling port end: (40±5)mm
8,Electricity source: AC 220V 50Hz
9, equipped with laser dust particle counter
a) Particle size measurement range: 0.3μm~25μm
b) Eight channels.
0.3μm, 0.5μm, 0.7μm, 1.0μm, 2.0μm, 3.0μm, 5.0μm, 10.0μm
c) Air flow: (28.3±1.4)L/min
d) sampling period: 1 second --- 59 minutes 59 seconds optional
e) printing function: built-in printer, can print the counting results
f) the maximum sampling concentration allowed: 35,000 / L
Comply with the standard.
YY T 0506.4-2106 "patient, medical personnel and equipment with surgical sheets, surgical gowns and clean clothing Part 4 dry state flocculation test method
YY/T 0506.2-2016 "Surgical sheets, gowns and gowns for patients, medical personnel and instruments Part 2: Performance requirements and test methods, Appendix A.2 Test methods for evaluation of cleanliness and particulate matter, Appendix A.4 Test methods for evaluation of flocculation".