Model:FTech-ISO1716
Standard:
• ISO 1716
Main Features:
1, The apparatus including: calorimeter apparatus, high-pressure oxygen bomb, cold water installations, computer and control system (which can record and display the curve), LCD screen and software;
Crucible made of stainless steel;
2, Box paint treatment; all internal parts are made of stainless steel, corrosion-resistant, easy to clean;
3, Ignition wire break is detection, and automatic ignition.
4, The test procedure automatically turn off.
5, Test abnormalities has sound alarm.
6, Computer control system:
7, Control box with USB / RS232 serial port with an external brand desktop or laptop operating control; High degree of automation, test procedure safe and reliable, easy to operate;
Specification:
1, Bomb capacity: (300 ± 50) mL.
2, Oxygen bomb pressure range: 0.5 ~ 3.5MPa, no leakage. The bomb pressure 21Mpa.
3, Timer: accuracy of 0.5s;
4. Maximum power: 0.5KW;
5, Ambient temperature: (10 ~ 35) ºC, humidity ≤80%;
6, Heat capacity of the instrument: about 10000J / K;
7, Heat repeatability error: ≤0.2%;
8, Temperature range: (4.5-50) ºC;
9, Temperature detection: 0.0001 ºC resolution, using the PT100 sensor;
10, Inner cylinder volume: 2000ml; Oxygen bomb Sealing: oxygenation pressure 2.5 ~ 3.5MPa, no leakage;
11, Motor power: 3W; stirring speed: the inner tube (375r / min);
Detailed Photos
After Sales Service
1. When get official order, equipment layout drawing and confirmation
2. Remote video training, installation guiding, trubleshooting by free.
3. On site training, installation guiding, all charges training fee+ traffic, hotel, board and lodging, Visa which provided by customer or paying.
4. 1 year warranty
5. Permanent software updated.