After-sales Service: | Yes |
---|---|
Warranty: | 1 Year |
Display: | Digital |
Control: | Manual Control |
Weight: | 0-100Kg |
Power Source: | AC220V |
Customization: |
---|
Suppliers with verified business licenses
Main purpose
This equipment is developed according to the new national standard GB24540-2009 "protective clothing acid and alkali chemicals protective clothing" Appendix D, mainly used to determine the liquid rejection efficiency of fabric acid and alkali chemicals protective clothing clothing material, so that the test liquid flows over the surface of the specimen, check whether the specimen is retained or penetrated, and calculate the liquid rejection efficiency.
Main technical indicators
1,Hard transparent tank: semi-cylindrical, inner diameter (125±5) mm, length (300±2) mm, inclined at 45°.
2,Syringe: specification (10±0.5) mL, needle hole diameter is (0.8±0.02) mm, the tip of the needle is flat.
3. Small beaker: 50mL capacity
4,Flat tip needle tip bottom from the bottom of the groove: (100±2)mm
5,Specimen size: (360±2)mm×(235±2)mm
6,Microcomputer automatic injection system injection speed: (10±1) s within continuous injection (10±0.5) mL liquid.
Conformity with standards
GB24540-2009 "Protective clothing for acid and alkali chemicals" Appendix D
Suppliers with verified business licenses