Product Description
Used for the melt mass flow rate (MFR) or melt volume flow rate (MVR) of a variety of plastic materials such as polyethylene, polybutylene, polystyrene, ABS, polyamide , fiber resin, acrylate, polyformaldehyde, fluoroplastics, polycarbonate to determine. The instrument adopts microprocessor control and integrated, digital measurement and control technology. It is programmed according to the standard test procedure. Fast temperature rise and high accuracy of constant temperature; after the packing, it can quickly restore the constant temperature. Chinese color crystal display, voice prompts, touch screen operation, convenient and easy to understand.
Reference Standards
GB/T3682-2000.
ISO1133-97, ASTM1238.
Technical parameters
Temperature control range: Room temperature~400 ºC.
Temperature control accuracy: ±0.5 ºC.
Display resolution: 0.1 ºC.
Diameter of barrel: φ9.550±0.025mm.
Barrel length: 152±0.1mm.
Diameter of piston rod head: φ9.475±0.015mm.
Length of piston rod head: 6.350±0.100mm.
Discharge port diameter: φ2.095±0.005mm.
Length of discharge port: 8.000±0.025mm.
Parameters of weights: 325g (including pressing rod), 875g, 1835g, 2500g, 2915g, 3475g, 4675g, 5000g, weights can be added according to user requirements.
Weights accuracy: ±0.5% .
Fluorine plastic is not available for the conventional model (optional).
Specifications
Dimension: 250mm×350mm×600mm.
Power supply: AC 220V 50/60Hz 3A.
Volume method: (customer optional).