Limited Oxygen Index Testing Machine (LOI) with England Imported Oxygen Sensor ISO 4589

Product Details
Customization: Available
After-sales Service: Yes
Warranty: 1 Year
Manufacturer/Factory & Trading Company
Gold Member Since 2017

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Plant Area
3074 square meters
Terms of Payment
LC, T/T
  • Limited Oxygen Index Testing Machine (LOI) with England Imported Oxygen Sensor ISO 4589
  • Limited Oxygen Index Testing Machine (LOI) with England Imported Oxygen Sensor ISO 4589
  • Limited Oxygen Index Testing Machine (LOI) with England Imported Oxygen Sensor ISO 4589
  • Limited Oxygen Index Testing Machine (LOI) with England Imported Oxygen Sensor ISO 4589
  • Limited Oxygen Index Testing Machine (LOI) with England Imported Oxygen Sensor ISO 4589
  • Limited Oxygen Index Testing Machine (LOI) with England Imported Oxygen Sensor ISO 4589
Find Similar Products

Basic Info.

Model NO.
FTech-ISO4589
Display
Digital
Control
Computer Control
Weight
0-100Kg
Power Source
AC220V
Oxygen Concentration Measurement Accurac
± 0.4%
Measurement Time Accuracy
± 0.5s
Measurement of Smoke Density Accuracy
± 3%
Transport Package
Plywood Packing
Specification
435(W)× 550(D)× 670(H)mm
Trademark
TESTech
Origin
China
HS Code
90248000
Production Capacity
100

Product Description

Introduction
TESTech Limit Oxygen Index Tester is designed according to ISO4589.1-2006, it applies to analysis the minimum oxygen concentration needed to maintain combustion of the sample when the sample is under under the test conditions specified in oxygen and nitrogen mixed gas;

This tester is installed with England imported Oxygen sensor, and automatic control the Oxygen and nitrogen flow with mass flow control meter. It is very easily and quickly to operate; Used for evaluating homogeneity of solidmaterials, laminates, foam, film and thin film in the specified conditions of flammability.
Model:FTech-ISO4589
Limited Oxygen Index Testing Machine (LOI) with England Imported Oxygen Sensor ISO 4589
Standard:
ISO4589.-2006
Technical Parameter:
1, Oxygen concentration measurement range: 0-100%;Accuracy of ± 2, Oxygen concentration 0.1%
2, Oxygen Mass flow meter range: 0~15.0L/min
3, Nitrogen Mass flow meter range: 0~15.0L/min
4, Response time: <10s
4,Oxygen concentration measurement accuracy: ± 0.4%     
5, Combustion tube top airflow 90mm ± 10MM/s
6,Combustion air flow in cylinder 6, 40mm/s ± 2mm/s, concentration control accuracy ± 0.1% 
7,Measurement time, accuracy ± 0.5S
8,Portable igniters, Jet flame down 16± 4mm adjustable;
9,Double flow meter and pressure gauge devices: precision: 1%
Features:    
1,England Imported ParaMagnetic oxygen sensor, accuracy is ± 0.1% ; without working life limit, it is definitely difference with electrochemistry oxygen sensor.  (It is difference with electrochemistry sensor, no need to change the sensor every year.)
2,Equipped with heat-resistant quartz glass tube, export ID 40mm;       
3,Equipped with supporting and not supporting sample holder   
4,Mass flow controllers, automatic regulating oxygen and nitrogen gas flowaccording to the set concentration;
5,Patent design of oxygen and nitrogen mixed unit to make sure the gas is mixed entirely
6,The expected oxygen concentration is set on screen, and PLC control system; No need manually operation;
7, LCD screen show test data such as flow rate, time, results in real time and oxygen concentration etc .
Control System:
1, PLC with touch screen system to make the test high automation;
2, After setting the oxygen concentrations, the system automatically adjustingthe oxygen concentration;
3, Key electrical components adopt imported brands, accurate, reliable, anddurable;
4, Equipped with a protection circuit;
Smoke Density System(Optional)
1, Optical absorption range: 0-100% can be measured continuously. Smoke density resolution 0.1%
2, Measurement of smoke density is accuracy: ± 3%;
3, Optical glass is easy to be cleaned, avoid to the negative of dust particles;
4, Light resource is 6V instrument special purpose incandescent lamps;
5, System is equipped with calibration system for light transmittance;

Specification

Model FTech-ISO4589
Size 435(W)×550(D)×670(H)mm
Power AC 220V, 50/60Hz, 5A
Weight Supplied
Operating manual With package
Exhuast 50 ℓ/s

 Limited Oxygen Index Testing Machine (LOI) with England Imported Oxygen Sensor ISO 4589
 Limited Oxygen Index Testing Machine (LOI) with England Imported Oxygen Sensor ISO 4589
 Limited Oxygen Index Testing Machine (LOI) with England Imported Oxygen Sensor ISO 4589
 

 
 

 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier