Product Description
TESTECH Oxyen Index Tester for the determination under the test conditions specified in oxygen and nitrogen gas mixture just to maintain a minimum sample combustion oxygen concentration required. This tester equipped with imported oxygen senser to make sure the measurement accuracy and durability. For the assessment of the performance of a homogeneous combustion of solid materials, laminates, foams, films and films under specified conditions.
Standards:ASTM D 2863, ISO 4589-2
Model:FTech-ASTM D2863C
Parameters:
1, The oxygen concentration measurement range: 0~100%;
2, The oxygen concentration in accuracy of ± 0.1%
3, N2 mass flow controller range:1.0~15.0L/min;Accuracy:2%
4, O2 mass flow controller range:1.0~15.0L/min;Accuracy:2%
5. Response time: <10s;
6.The gas flow rate in combustion column is 40mm / s ± 2mm/s;Concentration regulation accuracy ± 0.1%
7, The gas flow rate in top of the combustion column is 90mm ± 10mm/s;
- Portable igniter, downward spray 16 ± 4mm, flame adjustable range 10 ~ 30mm
9.The measurement time is 5 mins, and the accuracy is ± 0.5s
10, Double flow meter and pressure gauge device: accuracy ± 1%
11.Working pressure:0.1Mpa
12.Glass column inner diameter: 80mm , heigh 500mm.
Main Features:
- Equipped with a high temperature quartz glass column, the opening DIAM. of column cap is 40mm
- Equipped with samples holders for physically self-supporting specimens and supported film or sheet test specimens.
- Rotameter, regulating the flow of oxygen and nitrogen
- The patented gas flow mixing control unit is equipped with nitrogen and oxygen mass flow controllers to precisely control the gas flow.
- Double flow meter and pressure gauge device: accuracy ± 1%.
- Equipped with special igniter, easy to operate.
Control System:
1,The oxygen proportion is adjusted manually, and the oxygen concentration will be displayed automatically after stabilization.
After the oxygen concentration value is set, the system automatically adjusts the oxygen concentration;
2. The key electrical components are imported, accurate, reliable and durable;
Detailed Photos