Dry state microbial penetration test system consists of gas source generation system, testing body, control system, etc. It is used to conduct dry state microbial penetration test method on surgical sheet, surgical gown and clean clothes, etc. The dry state microbial penetration resistance tester/dry state bacteria resistance tester is used to determine the performance of material resistance to bacterial penetration on dry state particles within the size range of human dander, and the specimen is fixed on a container when conducting the test.
Equipment use.
Dry state microbial penetration test system by the gas source generation system, detection of the main body, control system and other parts, used for the surgical sheet, surgical gown and clean clothes, etc. dry state microbial penetration test method.
Conform to the standard.
YY/T0506.5-2009 "Surgical sheets, gowns and clean clothing for patients, medical personnel and instruments - Part 5: Dry state microbial penetration resistance test method
ISO 22612:2005 "Infectious agent protective clothing - Test method for penetration of microorganisms in the dry state".
Scope of application.
Dry state microbial penetration resistance tester / dry state bacteria resistance instrument is used to determine the material resistance to human dander size range of dry state particles on the performance of bacterial penetration, when conducting the test, the specimen is fixed on a container.
1,Vibration form: pneumatic ball shaker
2,Vibration frequency: 20,800 times / min
3,Vibration force: 650N
4, table size: 40cm × 40cm × 10mm
5, the table medium: marble plate
6,Experimental container: stainless steel experimental container
7,Working position: 6
8, equipped with biological safety cabinet
8.1, biological safety cabinet using four negative pressure, that is: the entire device of the left and right two sides, the rear, the bottom wall are negative pressure ducts, so that the work area and the external environment into a double layer of gas curtain and box isolation, while the work area is surrounded by negative pressure to ensure that samples do not leak.
8.2, biosafety performance: personnel safety, impact sampler microbial colony count ≤ 10 CFU / time, slit sampler colony count ≤ 5 CFU / time
8.3, airway tightness: the device's airway under the pressure of 500Pa for 30 minutes, the pressure drop ≤ 10%, and its penetration part in the soap bubble test conditions without bubbles.
8.4, cleanliness level: 100, both per IL air to ≥ 0.5u m particles of dust content ≤ 0.35
8.5, noise: the height of the table, 30cm from the front window, and 38cm upward, in the ambient noise of 55Db conditions, the arithmetic mean of four directions straight ≤ 58dB (A)
8.6, lighting: ≥ 680LUX front window glass: tempered, UV protection
8.7, UV intensity: ≥ 80uw/cm2
8.8, differential pressure display: (30 / 2) bit LCD display, the United States GE differential pressure sensor. LCD display shows the number indicates the pressure loss of the filter, its differential pressure sensor error ≤ 5%
8.9,Mechanical performance: The table surface of the equipment should be able to withstand 25Kg pressure without obvious sinking, bending and other deformations.
8.10, electrical strength: voltage resistance 1500V 1 minute without breakdown
8.11, liquid collection tank: used to collect the spilled liquid in the work area and equipped with a drain valve.
8.12, ultra-high efficiency filter is suitable for filtering airborne particles, smoke and microorganisms. The filtration efficiency of the imported ultra high efficiency filter (ULPA) is more than 99.999 % (0.3μm PAO method), so that the cleanliness of the working space can reach class 100. Equipped with dual-channel digital differential pressure display and electronic alarm system to display the pressure loss of exhaust and supply air filters in real time dynamically, the down air flow rate and inflow air flow rate in the operation area, dual detection, when the down air flow rate and inflow air flow rate fluctuate more than 20% of its nominal value, the sound and light alarm to indicate the down air flow rate and fluctuation. Front replacement, maintenance of filters and fans.
8.13,When the opening height of the front window of the safety cabinet exceeds the specified height of the front window operation opening, the audible and visual alarms are sounded and the interlocking system is activated. When the opening height back down to within the prescribed opening height, the alarm interlocking system is automatically lifted.
9,Working voltage: 220V 50Hz
1. When get official order, equipment layout drawing and confirmation
2. Remote video training, installation guiding, trubleshooting by free.
3. On site training, installation guiding, all charges training fee+ traffic, hotel, board and lodging, Visa which provided by customer or paying.
4. 1 year warranty